Enlarge
 
 

TkR-300CG HIGH-FREQENCY JETTING VENTILATOR

 Feature:

A. The device adopts impulse jet oxygen feeding technique with great flow rate of oxygen feeding in a shot time, when feeding oxygen, it adopts change mode major in reflux and many kinds of gas flow mechanism such as convection, laminar flow, sheet flow and so on concurrently, which is good to increase partial pressure of oxygen rapidly, At the same time fluctuation of chest caused by increasing oxygen flow into airway, it is good to discharge CO2.
B. Jet oxygen feeding is supported by pressure when inspiration, and guarantee certain PEEP effect when feeding oxygen, So as to reach good effectiveness of oxygen feeding.
C. Storage battery in the machine, it can be used while being charged, and can last for four hours after be charged.
D. Alarm with sound and light when power shortage or low airway pressure.
E. Digital display of import pressure sensor.
F. Three kinds of Alarm for upper-limitation of airway pressure for adult(2.0Kpa), child (1.0Kpa)and infant(0.5Kpa)
G. Pressure humidifier which has good humidigying effectiveness.

 Specifications:

 Frequency: 10~150 breaths/min
 I/E ratio: 1:1~ 1:3 and Inverse ratio
 Driving Pressure : 0~150KPa 
 Process Ventilation: T=10min, I/E=1:1.5
  ProI: f(150):f(35)=7:3
  ProII:  f(35):f(150)=7:3
 SIGH:6 breaths/min
 Alarm: Alarm with light and sound for no gas or power failure or airway pressure
   upper-limitation
 CPAP: 0.1~0.8KPa
 Maximum Oxygen Concentration: 85%
 Decompressor Limit pressure: 0.30MPa
 Inner Battery: 4Ah 12V
 Ventilation Modes:HFJV(High Frequency Jet Ventilation);CPAP(Continuous positive
  airway pressure); Process I and Process II ventilation; HFJV+SIGH;, Manual ventilation.

 
Copyrights ©JiangXi Teli Anaesthesia & Respiration Equipment CO.,LTD. GanICP0500248